Accurate measurement data are essential when validating thermal processes. Whether for sterilisation, cleaning or disinfection, you need reliable temperature and pressure measurements, even under extreme conditions. 

With Valilog, Normec offers its own data logger system for validation and process monitoring. The system has been specifically developed for use in reprocessing and validation procedures and supports organisations in collecting, analysing and documenting measurement data. 

Developed for the practice of validation 

Valilog was developed by experienced validation experts who deal with complex validation issues on a daily basis. As a result, the system is perfectly tailored to the requirements of practical application.

The system consists of: 

  • Temperature loggers with short sensors  
  • Temperature loggers with long sensors  
  • Dual temperature loggers  
  • Pressure loggers  
  • User-friendly validation software  

This combination enables a wide range of processes to be monitored and validated with precision.

Measure, calibrate and report 

Valilog provides a complete system for performing and documenting validation studies. We also provide both factory calibrations and DAkkS-accredited calibrations through our own calibration laboratory. 

This means you can: 

  • Measure temperature and pressure with a high degree of accuracy  
  • Work with equipment specifically developed for validation processes  
  • Benefit from calibration and support from a single partner  
  • Analyse measurement data easily using user-friendly software  
  • Rely on accurate documentation for audits and quality assurance  

Expertise of validation and calibration specialists

What sets Valilog apart is the combination of practical experience and technical expertise. The system has been developed by professionals who understand the challenges involved in validating reprocessing procedures. 

Our specialists not only help you select the right logger configuration, but also provide calibration, maintenance and software support. This helps you get the most out of your validation processes. 

Ready for future validation requirements 

Requirements for validation, quality assurance and process control continue to become more stringent. With a flexible and reliable data logger system, you have the measurement data you need to demonstrate that your processes remain under control. 

Valilog helps you measure accurately today while preparing for future standards, audits and quality requirements. 

Would you like to know more about Valilog?

Download our brochures

Calibration

We offer accredited calibrations and factory calibrations in our own calibration laboratory within 5 working days according to certified standards.

Scope of services of the permanent calibration laboratory

Measured variable / Calibration objectMeasurement range/ Measurement spanMeasurement conditionsExpanded measurement uncertainty ¹

Pressure

Absolute pressure pₐᵦₛ

0,01 bar to 7 bar

Direct comparison

Pressure medium: Gas

0,25 mbar

+ 4,5 -10-5 pₐᵦₛ

Temperature

Direct reading thermometers

Dataloggers

0 °C to 140 °CComparison with PT100 sensor with digital output in thermostated bath20 mK

¹ The smallest measurement uncertainly that can be stated is not realisable on every calibration object.

Flexible scope of accreditation

The calibration laboratory of Normec Valitech GmbH & Co. KG is accredited for a flexible scope.

Current list of standards and guidelines for accredited activities in the flexible scope:

Designation of the standard/directiveTitle of the standard directiveVersion
DKD-R 5-1Calibration of resistance thermometers09/2018 Rev. 0
DKD-r 6-1Calibration of pressure measuring devices03/2014 Rev. 3

Accreditations & Recognitions

We are accredited by the German Accreditation Body (DAkkS) as a manufacturer-independent testing laboratory. This certifies that we always perform our services in accordance with applicable standards and guidelines, and our QM system ensures consistently high quality. Here you can find our current DAkkS accreditation certificate , along with the detailed scope of application and a description of the testing procedures.

You might also be interested in these services

Validation of Cleaning and Sterilization Instructions

“A laboratory technician who performs medical tests using specialized equipment and ensures quality assurance.”

Validation of Medical Reprocessing Procedures

Frequently Asked Questions about valilog data loggers

Valilog is Normec Valitech's data logger system for monitoring and validating thermal processes in which temperature and pressure play an important role. 

Valilog is primarily used for the validation of reprocessing procedures, including cleaning, disinfection and sterilisation processes.

The system includes pressure loggers, temperature loggers with short and long sensors, and dual temperature loggers. 

Yes. Both factory calibrations and DAkkS-accredited calibrations are carried out in our own calibration laboratory. 

Yes. Valilog includes specially developed software for analysing, managing and documenting measurement data. 

Reliable measurement data are essential for demonstrating that processes consistently meet quality and validation requirements. 

Yes. The software supports structured recording of measurement data, making documentation readily available for audits and inspections.